What is the difference between fluorescence and photoluminescence




















Other types of luminescence, such as phosphorescence, occur much more slowly. Luminescent materials have many applications, and thus their development is an ongoing area of research in science. We use fluorescent light bulbs in the home, optical brighteners in our washing powder, luminescent materials in plasma TV screens, and luminescent inks in bank notes, stamps, cheques, and driving licenses, as anti-counterfeit measures.

To learn more about sensing genetic disorders with fluorescence and more see the chemBAM general experiments page. This work is licensed under a Creative Commons Attribution 4. Skip to content Home Definitions fluorescence vs luminescence. It is common to use two detectors and cross- correlate their outputs leading to a cross-correlation function that is similar to the auto correlation function but is free from after-pulsing when a photon emits two electronic pulses.

As mentioned earlier, when combined with analysis models, FCS data can be used to find diffusion coefficients, hydrodynamic radii, average concentrations, kinetic chemical reaction rates, and single-triplet dynamics. When particles pass through the observed volume and fluoresce, they can be described mathematically as point spread functions, with the point of the source of the light being the center of the particle.

A point spread function PSF is commonly described as an ellipsoid with measurements in the hundreds of nanometer range although not always the case depending on the particle. This Gaussian is assumed with the auto-correlation with changes being applied to the equation when necessary like the case of a triplet state, chemical relaxation, etc.

The expression is valid if the average number of particles, N, is low and if dark states can be ignored. FCS is often seen in the context of microscopy, being used in confocal microscopy and two-photon excitation microscopy. In both techniques, light is focused on a sample and fluorescence intensity fluctuations are measured and analyzed using temporal autocorrelation. The magnitude of the intensity of the fluorescence and the amount of fluctuation is related to the number of individual particles; there is an optimum measurement time when the particles are entering or exiting the observation volume.

When too many particles occupy the observed space, the overall fluctuations are small relative to the total signal and are difficult to resolve. On the other hand, if the time between molecules passing through the observed space is too long, running an experiment could take an unreasonable amount of time.

One of the applications of FCS is that it can be used to analyze the concentration of fluorescent molecules in solution. Here, FCS is used to analyze a very small space containing a small number of molecules and the motion of the fluorescence particles is observed. The fluorescence intensity fluctuates based on the number of particles present; therefore analysis can give the average number of particles present, the average diffusion time, concentration, and particle size.

This is useful because it can be done in vivo, allowing for the practical study of various parts of the cell. FCS is also a common technique in photo-physics, as it can be used to study triplet state formation and photo-bleaching. State formation refers to the transition between a singlet and a triplet state while photo-bleaching is when a fluorophore is photo-chemically altered such that it permanently looses its ability to fluoresce.

By far, the most popular application of FCS is its use in studying molecular binding and unbinding often, it is not a particular molecule that is of interest but, rather, the interaction of that molecule in a system.

By dye labeling a particular molecule in a system, FCS can be used to determine the kinetics of binding and unbinding particularly useful in the study of assays.

There are two types of luminescence: fluorescence and phosphorescence. It is a longer-lasting and less common type of luminescence, as it is a spin forbidden process, but it finds applications across numerous different fields. This module will cover the physical basis of phosphorescence, as well as instrumentation, sample preparation, limitations, and practical applications relating to molecular phosphorescence spectroscopy.

Phosphorescence is the emission of energy in the form of a photon after an electron has been excited due to radiation. In order to understand the cause of this emission, it is first important to consider the molecular electronic state of the sample. In the singlet molecular electronic state, all electron spins are paired, meaning that their spins are antiparallel to one another. When one paired electron is excited to a higher-energy state, it can either occupy an excited singlet state or an excited triplet state.

In an excited singlet state, the excited electron remains paired with the electron in the ground state. In the excited triplet state, however, the electron becomes unpaired with the electron in ground state and adopts a parallel spin.

When this spin conversion happens, the electron in the excited triplet state is said to be of a different multiplicity from the electron in the ground state. Electrons in the triplet excited state are spin-prohibited from returning to the singlet state because they are parallel to those in the ground state. In order to return to the ground state, they must undergo a spin conversion, which is not very probable, especially considering that there are many other means of releasing excess energy.

Because of the need for an internal spin conversion, phosphorescence lifetimes are much longer than those of other kinds of luminescence, lasting from 10 -4 to 10 4 seconds. Historically, phosphorescence and fluorescence were distinguished by the amount of time after the radiation source was removed that luminescence remained.

However, basing the difference between the two forms of luminescence purely on time proved to be a very unreliable metric. Fluorescence is now defined as occurring when decaying electrons have the same multiplicity as those of their ground state.

Because phosphorescence is unlikely and produces relatively weak emissions, samples using molecular phosphorescence spectroscopy must be very carefully prepared in order to maximize the observed phosphorescence. The most common method of phosphorescence sample preparation is to dissolve the sample in a solvent that will form a clear and colorless solid when cooled to 77 K, the temperature of liquid nitrogen.

Cryogenic conditions are usually used because, at low temperatures, there is little background interference from processes other than phosphorescence that contribute to loss of absorbed energy. Additionally, there is little interference from the solvent itself under cryogenic conditions. The solvent choice is especially important; in order to form a clear, colorless solid, the solvent must be of ultra-high purity. The polarity of the phosphorescent sample motivates the solvent choice.

Common solvents include ethanol for polar samples and EPA a mixture of diethyl ether, isopentane, and ethanol in a ratio for non-polar samples. Once a disk has been formed from the sample and solvent, it can be analyzed using a phosphoroscope.

While using a rigid medium is still the predominant choice for measuring phosphorescence, there have been recent advances in room temperature spectroscopy, which allows samples to be measured at warmer temperatures. Similar the sample preparation using a rigid medium for detection, the most important aspect is to maximize recorded phosphorescence by avoiding other forms of emission. Current methods for allowing good room detection of phosphorescence include absorbing the sample onto an external support and putting the sample into a molecular enclosure, both of which will protect the triplet state involved in phosphorescence.

Phosphorescence is recorded in two distinct methods, with the distinguishing feature between the two methods being whether or not the light source is steady or pulsed. When the light source is steady, a phosphoroscope, or an attachment to a fluorescence spectrometer, is used. There are two different kinds of phosphoroscopes: rotating disk phosphoroscopes and rotating can phosphoroscopes. After a light beam penetrates one of the disks, the sample is electronically excited by the light energy and can phosphoresce; a photomultiplier records the intensity of the phosphorescence.

The sample is placed on the outside edge of the can and, when light from the source is allowed to pass through the window, the sample is electronically excited and phosphoresces, and the intensity is again detected via photomultiplier.

One major advantage of the rotating can phosphoroscope over the rotating disk phosphoroscope is that, at high speeds, it can minimize other types of interferences such as fluorescence and Raman and Rayleigh scattering, the inelastic and elastic scattering of photons, respectively. The more modern, advanced measurement of phosphorescence uses pulsed-source time resolved spectrometry and can be measured on a luminescence spectrometer. The spectrometer employs a gated photomultiplier to measure the intensity of the phosphorescence.

The lifetime of the phosphorescence is able to be calculated from the slope of the decay of the sample after the peak intensity. The lifetime depends on many factors, including the wavelength of the incident radiation as well as properties arising from the sample and the solvent used. Although background fluorescence as well as Raman and Rayleigh scattering are still present in pulsed-time source resolved spectrometry, they are easily detected and removed from intensity versus time plots, allowing for the pure measurement of phosphorescence.

The biggest single limitation of molecular phosphorescence spectroscopy is the need for cryogenic conditions. This is a direct result of the unfavorable transition from an excited triplet state to a ground singlet state, which unlikely and therefore produces low-intensity, difficult to detect, long-lasting irradiation.

Because cooling phosphorescent samples reduces the chance of other irradiation processes, it is vital for current forms of phosphorescence spectroscopy, but this makes it somewhat impractical in settings outside of a specialized laboratory. However, the emergence and development of room temperature spectroscopy methods give rise to a whole new set of applications and make phosphorescence spectroscopy a more viable method.

Currently, phosphorescent materials have a variety of uses, and molecular phosphorescence spectrometry is applicable across many industries. Phosphorescent materials find use in radar screens, glow-in-the-dark toys, and in pigments, some of which are used to make highway signs visible to drivers. Molecular phosphorescence spectroscopy is currently in use in the pharmaceutical industry, where its high selectivity and lack of need for extensive separation or purification steps make it useful.

It also shows potential in forensic analysis because of the low sample volume requirement. Forms of Photoluminescence Resonant Radiation : In resonant radiation, a photon of a particular wavelength is absorbed and an equivalent photon is immediately emitted, through which no significant internal energy transitions of the chemical substrate between absorption and emission are involved and the process is usually of an order of 10 nanoseconds.

Fluorescence : When the chemical substrate undergoes internal energy transitions before relaxing to its ground state by emitting photons, some of the absorbed energy is dissipated so that the emitted light photons are of lower energy than those absorbed.

One of such most familiar phenomenon is fluorescence, which has a short lifetime 10 -8 to 10 -4 s. Phosphorescence : Phosphorescence is a radiational transition, in which the absorbed energy undergoes intersystem crossing into a state with a different spin multiplicity. The lifetime of phosphorescence is usually from 10 -4 - 10 -2 s, much longer than that of Fluorescence. Therefore, phosphorescence is even rarer than fluorescence, since a molecule in the triplet state has a good chance of undergoing intersystem crossing to ground state before phosphorescence can occur.

Relation between Absorption and Emission Spectra Fluorescence and phosphorescence come at lower energy than absorption the excitation energy.

Adapted from D. Freeman and Company, New York Adapted from C. Upper left: photoluminescence from quantum dot semiconductors Upper Right: chemiluminescence bioluminescence of jellyfish, Lower Left: radioluminescence of a tritium watch dial Credit: Nite Watches, www.

There are many types of luminescence which can be classified by the energy source which initiates the luminescence process. An overview of the various types of luminescence and their energy sources are given in Figure 2. Many of these luminescence processes have important scientific and industrial applications such as electroluminescence where light is emitted upon the recombination of electrons and holes after applying an electric field across a material, and is the operating principle behind light emitting diodes; and chemiluminescence where the light emission is initiated by a chemical reaction and used in biological assays and is responsible for the glowing of glow sticks.

However, the focus of this article is on photoluminescence which forms the basis of the powerful non-destructive spectroscopic technique, photoluminescence spectroscopy, that is used extensively in both academia and industry.

Figure 2: Types of luminescence and their energy sources. Photoluminescence is the emission of light from a material following the absorption of light. The word in itself is interesting in that it the combination of the Latin derived word luminescence and the Greek prefix, photo -, for light.

Any luminescence that is induced by the absorption of photons is called photoluminescence. This could equally be light emission from an organic dye molecule in solution Figure 3a , or band-to-band recombination of electrons and holes following photoexcitation of a semiconductor Figure 3b.

Figure 3: Examples of photoluminescence. Describing any photon absorption induced light emission as photoluminescence is accurate; however, it is common practice, particularly by chemists, to further subdivide photoluminescence into fluorescence and phosphorescence. It is often compared with a spinning top, either spinning in a clockwise or anti-clockwise direction.

However, this description is neither mathematically nor physically quite correct. In the Jablonski diagram for fluorescence see Fig. Within those states, there are several energy levels. The higher the level is, the more energy an electron possesses when being in that level. In the case of singlet states, the electrons have antiparallel spins. The electrons are lifted from the ground state S 0 , for example, to an energy level of the second excited state S 2 , when excited by electromagnetic radiation.

After excitation stops, the electrons only stay in that excited state for a short period of time ca. In doing so, energy initially can be released to the surroundings by vibrational relaxation. That means thermal energy is released by the motion of the atom or molecule until the lowest level of the second excited state is reached. The bigger gap between the second and first excited state is overcome by internal conversion. That describes an electronic transition between two states while the spin of electrons is maintained.

Now, the electrons can relax further due to more vibrational relaxation until they reach the lowest energy level of the S 1 state. Theoretically, the electrons could relax even further in a non-radiative way until they eventually reach the ground state again. However, it can be the case that the last amount of energy is too large to be released to the surroundings because the surrounding molecules cannot absorb this much energy.

Then, fluorescence occurs, which leads to an emission of photons possessing a certain wavelength. The emission lasts only until the electrons are back in the ground state. Since during all those transitions the electron spin is kept the same, they are described as spin-allowed [6,7,10].

For phosphorescence, things are a bit different see Fig. There are again an S 0 ground state and the two excited states, S 1 and S 2. Additionally, there is an excited triplet T 1 state which lies energetically between the S 0 and S 1 state. The electrons again have antiparallel spins in the ground state. Excitation happens in the same way as in fluorescence, namely through electromagnetic radiation.

The release of energy through vibrational relaxation and internal conversion while maintaining the same spin is the same here, as well, but only until the S 1 state is reached. Alongside the singlet states, a triplet state exists and so-called intersystem crossing ISC can occur since the T 1 state is energetically more favorable than the S 1 state. This crossing, like internal conversion, is an electronic transition between two excited states.

But contrary to internal conversion, ISC is associated with a spin reversal from singlet to triplet. This ISC process is described as "spin-forbidden".



0コメント

  • 1000 / 1000